| Hardware features |
| Tip configuration |
EVOlyzer 100: 2 or 4 tips; EVOlyzer 150: 4 or 8 tips; EVOlyzer 200: 8 tips
Various combinations of different tip types are possible |
| Tip types |
Teflon®-coated stainless steel tips and disposable tips with or without filter (200/1000 μl) |
| Syringe size |
1000 μl |
| Liquid handling features |
| Pipetting precision (CV) |
Volume |
Fixed tips |
Disposable tips 200 μl |
Disposable tips 1000 μl |
| Standard volumes * |
10 μl |
≤ 3.5% |
≤ 3.5% |
- |
| |
100 μl |
≤ 0.75% |
≤ 0.75% |
≤ 1% |
| |
* Typical results pipetting plasma in single pipetting mode (1000 μl syringe) |
| Liquid detection |
Down to 50 μl of conductive liquid can be detected in round-bottom microplates on standard
microplate carrier |
| Sample loading range |
Variable, starting with 16 sample tubes per grid up to 640 sample tubes (40 grids) depending on the worktable size |
| Sample input formats |
Sample tubes (tube diameter from approx. 10 up to 16 mm), microplates, strip plates or deep-well plates |
| Strip handling |
Up to 12 different assays per plate |
| Predilution options |
Common, serial or exclusive predilution, as well as several predilutions on one plate |
| Barcode identification |
Sample tubes, microplates, deep-well plates, strip plates, reagent containers |
| Supported formats |
Code 39, Code 128, Codabar 2 of 5 Interleaved |
| Barcode evaluation |
Reagent type, recording of lot number, lot expiry date |
| Incubation |
All incubators are suitable for dark incubation, and temperature is monitored |
| Ambient incubations |
Freedom EVOlyzer 100: 6 positions; Freedom EVOlyzer 150 & 200: 6/12 positions |
| Heated incubations |
Freedom EVOlyzer 100: 6 positions; Freedom EVOlyzer 150 & 200: 6/12 positions with or
without shaking |
| Temperature range |
RT-incubator: ambient temperature measured with accuracy of ± 1.5°C |
| |
Heated incubator MIO-2™: from 5°C above ambient temperature up to 46 °C. Incubator is calibrated at 37°C or 46 °C with accuracy of +/- 0.5°C. The temperature accuracy over the plate +/- 1.0°C |
| Washer: HydroFlex™ |
| Manifold |
8-way manifold |
| Wash channels |
4 channels for wash buffers, wash buffer capacities: 2 x 2 l and 2 x 4 l with liquid level sensing |
| Wash modes |
Strip or plate mode |
| Photometer: Sunrise™ |
| Measurement range |
340 - 3 99 nm: 0 - 3 .0 OD; 400 - 750 nm: 0 - 4 .0 OD |
| Wavelength selection |
4 or 6 fixed interference filters (340 – 750 nm) or gradient filter (400 – 750 nm) |
| Measurement time |
Single wavelength: 6 seconds; dual wavelength: 8 seconds |
| Software |
Supporting 21 CFR part 11, operating system Windows® XP SP2 |
| Interface |
Bidirectional ASTM* (NCCLS** LIS1*** /NCCLS LIS2) interface |
| Dimensions |
Freedom
EVOlyzer 100 |
Freedom
EVOlyzer 150 |
Freedom
EVOlyzer 200 |
| Height |
910 mm/35.8" |
910 mm/35.8" |
910 mm/35.8" |
| Width |
1520 mm/59.8" |
1890 mm/74.4" |
2490 mm/98" |
| Depth |
800 mm/31.5" |
800 mm/31.5" |
800 mm/31.5" |
| Weight |
210 kg |
260 kg |
290 kg |
| Power supply requirements |
Power: 600 VA (EVOlyzer 100) and 1200 VA (EVOlyzer 150 & 200) Voltage:
100 - 120 VAC and 220 - 24 0 VAC, frequency 50/60 Hz |
| Operating conditions |
Temperature: 15 - 32 °C/59 - 90°F; relative humidity: 30 - 80% (non condensing) at 30°C/86°F |
| Compliance |
CE marked; compliant with the directive of in vitro diagnostic medical devices 98/79/EC |